Fwd: Fwd: Agilent 82350b can't recover from read timeout?

Dirk Niggemann <[email protected]>
Newsgroups gmane.linux.hardware.gpib.general
Message-ID <CAKrfojgY+Efzqo4e3vOEoC1vU0dEPnV_BrdDVuG1PJ+CHge_tw@mail.gmail.com>
H Frank,

I checked the value returned from the DIR register and it's almost always
0xa so the last byte sent on the previously successful read. Or 0 if we
never read anything.

I am already testing with the _unaccel driver because i wanted to make sure
this wasn't a side effect of using the FIFOs- sorry, I should have
mentioned that.

Regards,

Dirk

On Wed, Jul 31, 2019 at 5:41 PM Frank Mori Hess <[email protected]> wrote:

> On Tue, Jul 30, 2019 at 4:14 AM Dirk Niggemann <[email protected]>
> wrote:
> >
> > Yes, that works too.
> >
> > That leaves the question of why there is a byte to read when nothing was
> sent?
> >
>
> Oh, maybe there isn't a byte.  Reading the DIR register tickles the
> data holdoff logic, so maybe that could cause it.  Another thing you
> could try is using "agilent_82350b_unaccel" as the board type, to take
> the fifos out of the picture.  Then the driver will only use the
> tms9914 directly.
>
>
> _______________________________________________
> Linux-gpib-general mailing list
> [email protected]
> https://lists.sourceforge.net/lists/listinfo/linux-gpib-general
>

_______________________________________________
Linux-gpib-general mailing list
[email protected]
https://lists.sourceforge.net/lists/listinfo/linux-gpib-general
lmpx.com only provides a reader for public news (NNTP) servers. It is not affiliated with the servers or forums shown here and is not responsible for the content of articles, which is written by their respective authors.