Re: Fwd: Fwd: Agilent 82350b can't recover from read timeout?

Frank Mori Hess <[email protected]>
Newsgroups gmane.linux.hardware.gpib.general
Message-ID <CAJz5OpcYkOjVB7+BW42PFzDVZgkCa=1G1TG-86Mx4EqX+x2vUQ@mail.gmail.com>
On Thu, Aug 1, 2019, 02:23 Dirk Niggemann <[email protected]> wrote:

> Hi Frank,
>
> Yes, disabling TCS works perfectly, as you suggested.
> Do you think that's the correct fix for this? Or should we aim for
> something like a read of DIR register just before every TCA
>


Yes disabling tcs is best.  The rhld command or read of DIR causing
transitions to NRFD is actually nonsensical behavior.  There are several
errors in the acceptor state machine described by the tms9914a manual.  My
guess is the designers of the 82350b copied mistakes in the tms9914a
documentation into their hardware.


>>