Re: Agilent 82357B repeatable hard failure
dave penkler <[email protected]> Sat, 30 Sep 2023 10:03:09 +0200
| Newsgroups | gmane.linux.hardware.gpib.general |
|---|---|
| Message-ID | <CAL=kjP3dJcNmjHLWxZQ2CM4iccZNw4sHeOSus2Lxff+tPymOAw@mail.gmail.com> |
Hi Jim, On Fri, 29 Sept 2023 at 14:46, Jim Houston <[email protected]> wrote: > Hi Dave, > > Thanks again for all of your efforts in debugging this problem. I have > enjoyed working with you. > I enjoyed working with you on this. Your logic analyser trace at the interface level of the 9914 IC was a huge help. > I like setting ATN from the end of the read function. I think this > works well for > the normal case where we send a query and get a reply. > I would avoid duplicating code and call the agilent_82357_take_control > function. > > OK, I had done an in-line write to the auxiliary command register to avoid the reading of the status byte in the timeout loop. But it can't hurt so I'll go with your suggestion. I see a balance between the chance that setting ATN might break > something against > knowing that having a correct ibsta() value will fix something. > > Agreed. > Jim Houston > -Dave _______________________________________________ Linux-gpib-general mailing list [email protected] https://lists.sourceforge.net/lists/listinfo/linux-gpib-general