Re: Agilent 82357B repeatable hard failure
Jim Houston <[email protected]> Tue, 3 Oct 2023 10:00:50 -0400
| Newsgroups | gmane.linux.hardware.gpib.general |
|---|---|
| Message-ID | <[email protected]> |
Hi Dave, Everyone, I had a look at the patch and it looks good. I also did a quick test and it works properly. The patch is against an intermediate version so I had to do a small fix to make it apply to SVN. JIm On 10/2/23 08:31, dave penkler wrote: > Hi Jim, > I refactored the take control function to avoid reading the status and > checking ATN when called from read. > See the attached patch. > Would this work for you ? > -Dave > > On Sun, 1 Oct 2023 at 14:48, Jim Houston <[email protected]> wrote: > > Hi Dave, Everyone, > > The loop reading the status and checking for ATN in > agilent_82357a_take_control > was helpful in finding the problems with the ADSR but maybe it > isn't needed anymore. > I did a couple tests. I tried setting timeout to 1 to check that > ATN is there on the first read. > Then I took out the loop and the timeout check. I ran the > hp3478_read_calibration script > in a loop for a few minutes. It works. > > I checked the times for a hp3478_read_calibration run and removing > the check > for the ATN reduced the time from 7.6 seconds to 7.3 seconds. > > JIm > > _______________________________________________ Linux-gpib-general mailing list [email protected] https://lists.sourceforge.net/lists/listinfo/linux-gpib-general